Setting Up Ports and Channels
120
How to Set Up a Port Or Channel
for Signals To Be Analyzed
To set the parameters for a port or terminal that returns signals or for a
channel that receives signals:
1 Open the Parameter Editor for the port or channel (see "How to Start
the Parameter Editor for Setting Port Parameters" on page 107 or
"How to Start the Parameter Editor for Setting Channel Parameters"
on page 108.
Figure 50
Timing Parameters for a Data Analyzer
Note that the window displays the identification of the frontend.

How to Set Analyzer Timing Parameters

1 Check the units.
The Parameter Editor displays default units and has default vernier
steps. Both can be changed (see "How to Change Units and/or Vernier
Steps" on page 69).
2 Set the timing options:
– Delay: The total delay for capturing received data is composed of a
relative delay (in fractions of system clock cycles) and an absolute
delay (independent of the system clock).
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1

How to Set Up a Port Or Channel for Signals To Be Analyzed