Introduction to the System
40
On a module with analyzer frontends, all connectors can either be in Pause
mode or they must all be in another mode than Pause, for example,
Capture or Acquire.
Data to Connector Assignment
The algorithm how the available segment data is assigned to the
connectors is as follows:
• The first terminal within a port gets trace0, the second gets trace1 and
so on.
The assignment to the connectors depends on what connections you
have selected from the terminals to the connectors in the Connection
Editor.
• If a terminal is connected to a connector where channels are added, the
connector that holds the connection gets the first trace. An added
channel gets the next trace.
In this case an exception to the rule "from top to bottom" is made.
Added channels are assigned from bottom to top.
Loops
Loops are used to repeat data blocks.
Looping Example
A sequence to loop infinitely 1 Kbit portions of a 2
64 bits might look as follows:
Figure 9
Simple Test Sequence With Loops
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
Data Generation Principles
15
–1 PRBS and a pause of