5 Maintenance and troubleshooting

Troubleshooting error messages

Problem
Cause
The robot does
The robot teachpoints might be set
not place or pick
incorrectly, the Labware MiniHub
labware correctly
has moved after the teachpoints
at the Labware
were set, or the shelves are bent.
MiniHub.
The Labware
The motor is disabled, possibly
MiniHub has
because an obstacle prevented the
stopped and the
Labware MiniHub from rotating.
motor status light
on the base is
red.
Related information
Troubleshooting error messages
About this topic
Communication errors
58
For information about...
Labware MiniHub components
Hardware problems
Reporting problems
The following tables list commonly encountered error messages, the causes of
the errors, and ways to resolve the errors.
If you are still experiencing problems with the Labware MiniHub after trying
the solutions, or if an error not on the list is displayed, contact Agilent
Automation Solutions Technical Support.
For protocol- related errors, see the applicable automation control software
user guide, such as the
The following error messages are caused by communication problems:
The command timed out.
Attempted to communicate without being connected, please connect and
retry the operation.
Failed to connect.
TCP connection has not been established.
Solution
Check and reset the robot teachpoints.
See
"Setting robot teachpoints" on
page
37.
Check for and replace bent shelves.
Remove the obstacle, click Retry, and
resume operation.
Reinitialize the Labware MiniHub. If the
problem persists, power cycle the device.
See...
"Hardware overview" on page 4
"Troubleshooting hardware problems"
on page 56
"Reporting problems" on page 60
VWorks Automation Control User
Guide.