10.9 Replacing power supply cable

1. Switch off the X-ray unit and disconnect the power supply cable from mains.
2. Remove the generator housing as described in section 10.1 "Removing generator
housing" on page 45.
3. Disconnect the power supply cable from the Generator PCB. Unscrew the power supply
cable strain relief and remove the cable.
4. Use the correct power supply cable: for mains voltage 100 V~/110-115 V~ power supply
cable Planmeca part number is 10006683 and for 220-240 V~ 10006584.
5. Check that the wires extend from the outer housing as shown below (1). If not, peel the
outer housing.
6. Peel 8 mm housing from the lead ends (2).
1
7. Route the power supply cable through the frame opening and attach the strain relief.
Tighten the strain relief as described in section 6.4 "Power supply cable" on page 31.
56
Planmeca ProX
70 mm
Yellow/ Green
Brown
Blue
30 mm
2
8mm
3000 mm