Connecting The Power Cable - Planmeca ProOne Manual

Hide thumbs Also See for ProOne:
Table of Contents

Advertisement

7.5 Connecting the power cable

7.5.1 Permanently installed X-ray units (power cable without plug)
WARNING
Ensure that the wires in the wall opening are not connected to live voltage.
1.
2.
3.
4.
5.
Installation Manual
Remove a knockout in the junction box.
Attach the strain relief to the power cable.
If needed, shorten the length of the power cable.
Route the wires of the power cable into the junction box.
Attach the junction box to the wall according to the local regulations.
7 INSTALLING THE X-RAY UNIT
Planmeca ProOne
25

Advertisement

Table of Contents
loading

Table of Contents