Guidance and manufacturer's declaration - electromagnetic immunity
Planmeca ProX intraoral X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of Planmeca ProX intraoral X-ray unit should assure that it is used in
such an environment.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which Planmeca ProX intraoral X-ray unit is used exceeds the
applicable RF compliance level above, Planmeca ProX intraoral X-ray unit should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating Planmeca ProX X-ray unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Technical manual
IEC 60601
Compliance
test level
3 Vrms
3 V/m
Electromagnetic environment-
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the Planmeca ProX X-ray unit,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d
d
d
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
3 EMC INFORMATION
guidance
=
1,2 P
80 MHz to 800 MHz
=
1,2 P
800 MHz to 2.5 GHz
=
2,3 P
should
be
less
Planmeca ProX
than
the
b
17