3.4 Setting the Output Condition in the FG Mode

Selecting the Output Channels and Range
Selecting the Output Waveform
IM 707281-01E
In this section, you will be using the operation panel that appears by double-clicking the
module icon in the station list window as shown in the figure below.
Turn ON/OFF the output
Select the channel to output
Select the output waveform
Select the output range
Set output frequency
Set the amplitude of the ouput voltage
Set the phase
Set the offset voltage
Select the channel for which to set the output in the [CH] list box. Clicking the [On] box
enables the output. Select the range from [1 V], [2 V], [5 V], and [10 V] in the [Range] list
box.
Select the output waveform in the [Func] list box. The choices are shown below. The
default is [Sine].
• Sine
: Sine wave
• Pulse
: Pulse wave. Set the duty cycle from 0 to 100% (resolution: 0.1%).
• Ramp
: Ramp wave
• Triangle : Triangular wave. Fixed to 50% symmetry.
• Arbitrary : Arbitrary wave
• DC
: DC output (Set the output voltage in [Offset].)
Setting waveform inversion
Check the [Invert] box to reverse the positive and negative polarities of the selected
output waveform.
Invert the waveform
Select the output mode
Set the duty cycle, valid
only when the output waveform
is set to [Pulse]
Turn ON/OFF sweep
Execute manual trigger
Turn ON/OFF the waveform
output
Load arbitrary waveform data