3.4 Setting the Output Condition in the FG Mode
Selecting the Output Mode
Select the output mode with the [Output Mode] option buttons. The choices are shown
below. The default is [Cont].
Cont: Continuous oscillation
Continuously outputs the waveform from the time the output is turned ON by clicking the
[Output] button until the output is turned OFF by clicking the [Output] button again.
Trigger: Trigger oscillation
After the [Output] button is clicked to turn ON the output, the waveform is continuously
3
output when a trigger occurs. This continues until the output is turned OFF by again
clicking the [Output] button.
Gate: Gate oscillation
After the [Output] button is clicked to turn ON the output, the oscillation starts when the
gate is enabled. When the gate is disabled the oscillation stops. The waveform output
unit is one cycle.
Note
When the output is OFF, the waveform output terminal of the module outputs 0 V.