Datcon DT4120 Series Operating Instructions Manual page 38

Temperature / resistance / potentiometer meter-transmitter
Table of Contents

Advertisement

38
11. After entering the correction value press MENU button.

You will see:
12. For entering the second correction value press▲button.

You will see:
by pressing the ENTER button.
13. According with steps 5.-10. enter the next correction
values using the same procedure:
temperature: 100.0 °C, resistance: 0.02 Ohm.
14. After entering the correction value press MENU button.

You will see:
15. For entering the third correction value press▲button.

You will see:
by pressing the ENTER button.
16. According with steps 5.-10. enter the next correction
values using the same procedure:
temperature: 130.0 °C, resistance: -0.08 Ohm.
17. After entering the correction value press MENU button.

You will see:
(18. If you don't want to exit the menu (you want to check or
you want do further settings), then you may select the
desired menu items by pressing the ▲ button:


,
Note:
You can enter the correction values:


,
example:
 
on the display.
on the display. Enter the menu item
on the display.
on the display. Enter the menu item
on the display.

,

,
in an optional sequence, for

,
DT4120 xx xx (PS)
20170324-V0