Datcon DT4120 Series Operating Instructions Manual page 37

Temperature / resistance / potentiometer meter-transmitter
Table of Contents

Advertisement

DT4120 xx xx (PS)
20170324-V0
3. Enter the menu item by pressing the ENTER button.
4. You see the mnemonic:
The digit "1" is blinking indicating that it can be modify.
It's meaning is: first correction point.
Enter the menu item by pressing the ENTER button.
5. You see on the display:
The left digit (
) is blinking indicating that it can be modify.
6. Pressing the ► button you can select the next digit.
7. Pressing the ▲ button you can increase the value of the
blinking digit:




,
,
,
,
8. When the left digit is selected, pressing ▲ button, the
sequence is:




,
,
,
,
So you can enter the desired temperature value between
-200.0 °C - 800.0 °C.
9. Type-in the first temperature value taken from the table

above: 70.0 °C:
10. After entering the temperature value press MENU
button, than you can type-in the correction resistance value
(Rn - Rm) in two decimal format.
Factory default value is: 0.00 Ohm:
You can modify this value to the value taken from the table
above. The range of the correction resistance value is
between: -2.00 Ohm-2.00 Ohm.
This value is in this example 0.32 Ohm:







,
,
,
,
,





,
,
,
,
,
.
on the display.
(Factory default).

,
, etc.


,
,
, etc.

.

.
37

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dt4120 ps series

Table of Contents