4.2.3
Removing the liner protectors
NOTICE
Sensors with PTFE or PFA liners only!
The following information in only relevant for sensor types with PTFE or PFA liners!
Sensors with PTFE liner
At delivery the sensor is mounted with wooden blanks to hold the liner in place during
transportation and storage. After only a few hours without the blanks, the liner will creep back
towards its original shape and installation will be more difficult to carry out.
Figure 4-6
• Remove the blanks immediately before mounting the sensor
Figure 4-7
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
Sensor with wooden blanks
Removing wooden blanks
Installing/mounting
4.2 Sensor installation
23