Dynamic Host Configuration Protocol (DHCP)

Setup (DHCP)

Software required
The modules in the stack must have at least firmware from set A.06.34 and the above mentioned
modules B.06.40 or above (must from the same firmware set).
1 Note the MAC address of the LAN interface (provided with G1369C LAN
2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Table 24
SW 4
ON
232
Interface Card or Main Board). This MAC address is on a label on the card
or at the rear of the main board, e.g. 0030d3177321.
On the Instant Pilot the MAC address can be found under Details in the LAN
section.
Figure 63
LAN Setting on Instant Pilot
Card or the main board of above mentioned modules.
G1369C LAN Interface Card (configuration switch on the card)
SW 5
SW 6
OFF
OFF
SW 7
SW 8
Initialization Mode
OFF
OFF
DHCP