General Error Messages

General error messages are generic to all Agilent series HPLC modules and
may show up on other modules as well.

Timeout

Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2
Agilent 1200 Infinity Series DAD User Manual
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Error Information

General Error Messages

Suggested actions
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.