Spectrum Settings

Store
None
Apex
Apex + Baselines
Apex + Slopes
Apex + Slopes + Baselines
All in Peak
Every 2nd spectrum
Range
Agilent 1200 Infinity Series DAD User Manual
Setting up the Detector with Agilent ChemStation
Defines at which points on "signal A" spectra will be
taken and saved. Signal A is used to control the "peak
controlled spectra acquisition"; the other signals have no
influence on spectra acquisition.
Limits:
190.0 to 640.0 nm in steps of 0.1 nm for both low and high
values. The high value must be greater than the low value
by at least 0.1 nm.
No spectra are taken.
Spectra are taken at the apex of the peak.
Spectra are taken at the apex and baselines of the peak.
Spectra are taken at the apex, upslope, and down slope of
the peak.
Spectra are taken at the apex, baselines, upslope, and
down slope of the peak.
All spectra within the peak are taken.
Spectra are taken continuously as for All, but only every
second spectrum is stored; other spectra are discarded.
This reduces the amount of data storage necessary.
Range defines the wavelength range for spectral storage.
Limits: 190 to 640 nm in steps of 1 nm for both low and
high values. The high value must be greater than the low
value by at least 2 nm.
4
Using the Module
61