Dark Current Test

The dark-current test measures the leakage current from each diode. The test
is used to check for leaking diodes which may cause non-linearity at specific
wavelengths. During the test, the slit assembly moves to the dark position,
cutting off all light falling onto the diode array. Next, the leakage current from
each diode is measured, and displayed graphically. The leakage current
(represented in counts) for each diode should fall within the limits.
When
In case of problem.
1 Run the Dark Current Test with the recommended user interface (for further
Agilent 1200 Infinity Series DAD User Manual
information see Online-Help of user interface).
Figure 49
Dark Current Test – Results
Test Functions and Calibration

Dark Current Test