5.6 Configuring the Hardware

Before the Ultium Motion System can be used, the device software settings must be configured to
recognize the different components that make up the system. Follow the below instructions to update
the receiver firmware, sensor firmware, and populate sensors to prepare for a data collection.
Step 1
Open MR3, typically listed under Noraxon ->
MR3
Click on the Hardware Setup button in the upper
right-hand corner.
Step 2
Select the Ultium icon, within the 'New Device'
column, and click on the Insert button.
Note: Make sure the Receiver is attached to the
USB port of the computer.
Step 3
The Ultium Settings dialog will appear as shown.
Within the "Settings" tab, select the desired
collection Sampling Rate.
Select a RF Network from the RF Network list. In
most cases the default "1" will work. However, if
using multiple Ultium systems they must be
placed on separate networks.
If using MyoSync, check the Use Noraxon
MyoSync checkbox.
Note: Firmware updates for the Receiver will be
indicated here, if they are required.