7.2 myoMOTION Software Settings

myoMOTION Software settings are in a myoMOTION tab in the Software Setup in the Home screen.
There are 3 different options: Magnetometer Control, Magnetometer Table, and Secondary Knee
Angles.
7.2.1 Magnetometer Control
Selecting this feature in the Software setup allows the user the option to turn off the magnetometer in
the Measure screen after a calibration (the magnetometer is needed for a clean calibration) before
taking a record. Only users with a high amount of magnetic interference that is making data unusable
should utilize this feature. Not using the magnetometers while recording data can cause drift over the
course of the record, so using this feature only for short records is advisable.
After calibrating myoMOTION, on the right-hand
side of the Measure window, you will see an
option that says "Magnetometer" with a Yes/No
option. Click No to disable the magnetometers.
The option to disable magnetic warnings also
present. Magnetic warnings is the indicator that
shows you a red magnetic value in the
magnetometer table if the magnetic interference
is too high.
Note: Magnetic warnings will not work if the
magnetometer is set to "off" within Hardware
settings.