7.5 Anti-Wobble Mode
Anti-wobble mode can be used to smooth data susceptible to soft tissue artifact. Soft tissue artifact
occurs when the sensor moves relative to the skin, which creates a "wobble" that is not indicative of
normal human movement.
Step 1
Open the recording of interest from the
Database screen.
Step 2
Within the record Viewer screen, select the
model settings button that looks like a note pad
at the bottom of the avatar window. This opens
the Model settings tool bar.
Step 3
Select the desired Anti-wobble smoothing
option.
1. Light (BW 15 Hz on 300 ms residual)
2. Medium (BW 10 Hz on 300 ms residual)
3. Heavy (BW 5 Hz on 300 ms residual)