The lamp inside the measuring head may be replaced by a technician. Please
follow the instructions below:
Tools and parts needed:
1 Allen key 3 mm
1 Allen key 2.5 mm
1 Allen key 2 mm
1 Spare lamp
Note: The spare lamps are aligned optically and
then glued to the lamp holder part. This
alignment process may only be performed at
Proceq SA Factory.
Open the measuring head housing:
1.
Loosen the three screws on the right side of
the cable on the back side (Allen key 2
mm).
2.
Loosen the three central screws on the
front side (Allen key 3mm).
3.
Loosen the four screws on the cooling air
outlet on the front side (Allen key 2.5mm).
Page 61