Instructions for Use
(continued)
PC Program Config
(continued)
Measurements with Show All Values
If the check-mark at Show All Values is activated, all values the sample are displayed. If you click the last
column called Description you can add more detailed information to this specific value. The last mea-
sured value is always shown and the sidebar is automatically scrolling down if you activate the check-box
Show Actual Measurement.
Figure IS-11, Measurements with Show all Values
Load Options
To load the instrument configuration from the AR5 refractometer to the PC, press LOAD under the COM-
port selector. After the successful download of the settings, the serial number and the firmware version of
the device will be displayed.
By using Edit you can take a look at the different settings and modify them if you want to. It is recommend-
ed to save the "original" configuration on your PC before changing the settings of the instrument. In the
future you can restore previous settings with this saved configuration.
Figure IS-12, Edit Refractometer Settings
26
1400700-101 Rev. A