Analyzer Customer-Supplied Mounting Plate with Pipe
Attached Pipe Nipple Installation
If you can't mount the sensor to the process directly because the wall is curved
sharply, covered with insulation, or obstructed with reinforcing members, you
can instead weld the AMETEK-supplied mounting plate with attached pipe
nipple to a your pipe nipple.
1. Weld a short 3-1/2" diameter pipe nipple to the duct and extend through
2. Insert the AMETEK-supplied pipe nipple (on the end of the mounting
3. Weld the mounting plate to your pipe nipple.
4. Wrap the exposed portion of your pipe nipple with a minimum of 1" thick
5. Mount the sensor to the process as described in the "Mounting the Sensor
Figure 9.
WDG-V HP Series Analyzer
customer-supplied mounting
plate with pipe nipple
installation.
40 | Thermox
®
WDG-V HP Series and WDG-V/S Series Analyzers
the wall. The shorter the pipe nipple, the better.
plate) into your pipe nipple. If the pipe nipple extends past the inside of
the refractory wall on a process which has a gas temperature over 650 °C,
cut down the nipple before installation so it is flush with the inside wall.
weatherproof insulation and heat trace it. This will keep the sample gas
above its dewpoint to prevent sample condensation that can plug sensor
plumbing.
to the Process" section.
for WDG-V HP Series