Instructions for Use
(continued)
PC Program Config
(continued)

Successful Connection

If the device is connected correctly the MEASURE and TEST functions will be displayed and can be ac-
tivated.
Figure IS-05, Loading Successful - Choose Measure

Measure

If you click MEASURE, the window shown in Figure IS-06 opens. Here you have the opportunity to adjust
the frequency of measurement in the Sample Interval field. For example, if you type 5, it means that the
refractometer measures and displays the results every 5 seconds. To start the measurement just click
START. After you complete the measuring process by activating STOP, you can save the displayed val-
ues in a csv format with SAVE. If you click CLEAR, the table with your measurements will be permanently
deleted.
Your device is equipped with a built-in peltier thermostat, so you can easily re-adjust the measuring tem-
perature by entering the desired temperature in the SET TEMP field and then selecting SET TEMP.
Figure IS-06, Measurements without Show all Values
-continued-