Instructions for Use
(continued)
PC Program Config
(continued)
Calibrate
To activate this button, you first have to load the instrument configuration in the start menu. Then you have
the opportunity to execute a 1 POINT- or 2-POINT calibration. The values of the Slope and the Offset can
be modified here.
To operate a 1 POINT calibration the Brix value has to be entered and confirmed by activating the STORE
button. Now you can see the measured and the expected value for the 1 POINT. sample.
To perform a 2 POINT calibration choose 2 POINT and type in a second Brix value in the field for the
second point.
Figure IS-08, Calibrate
-continued-
24
1400700-101 Rev. A