Function And Safety Checks; Safety Checks; General; Maintenance - Canon OMNERA 400T Operator's Manual

Manual-positioning digital radiographic system
Hide thumbs Also See for OMNERA 400T:
Table of Contents

Advertisement

7 Function and Safety Checks

7.1 Safety checks

7.1.1 General

Note!
Before performing any maintenance please read the safety chapter.
Note!
For exchange of the collimator light field lamp, see the Collimator manual.
If any malfunction is detected, the entire equipment must be taken out of use until the mal-
function is eliminated by a service engineer from the supplier or by the local technical staff
trained by the supplier.
Daily and monthly checks are normally performed by the user/operator.
Annual checks shall be performed either by local technical staff trained by the supplied or au-
thorized service representatives.
The Manufacturer recommends use of the checklist, Appendix B.

7.1.2 Maintenance

To ensure the safety of the patients, the operators, and third parties, and to maintain reliabil-
ity, the equipment must be checked according to Function and Safety Check list. If national
rules or regulations specify more frequent checks and/or maintenance, such regulations
must be observed.

7.1.3 Scheduled checks for the user

Item
Cleaning of parts in contact with the patient
Cleaning of main body of the equipment
Other maintenance and items as follows should only be performed by trained service
personnel.
Item
Checking for abnormal sounds
Checking for fallen foreign matter
Checking for oil leaks, etc
Checking the emergency stop buttons
Operator´s Manual
2020-09-18
Function and Safety Checks
Inspection period
Every day
Every 1 month
0180-095-418-EN Rev. 6.1
Page 129
Safety checks
OMNERA®