2.5 Installation and repair

WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
WARNING!
No modification of this equipment is allowed.
CAUTION!
Only service technicians are allowed to open the covers.
CAUTION!
Do not remove, disassemble, change, modify, repair, or add any part.
CAUTION!
When installing this equipment in a different location, contact the manufacturer or the
designated dealer.
Modifications of, or additions to, the product must be made in accordance with the legal regu-
lations and generally accepted engineering standards.
The manufacturer cannot assume responsibility for the safety features and for the reliability
and performance of the equipment, if:
• installation of equipment expansions or modification are not approved by the
manufacturer.
• installation of equipment expansions or modification are not carried out by persons author-
ized by the manufacturer.
• components are not replaced by original spare parts in case of a malfunction.
• the electrical installation of the room concerned does not meet the requirements or the cor-
responding national regulations.
• the product is not used in accordance with the operating instructions.