Operating The Wall Stand; General - Canon OMNERA 400T Operator's Manual

Manual-positioning digital radiographic system
Hide thumbs Also See for OMNERA 400T:
Table of Contents

Advertisement

4.5 Operating the Wall Stand

4.5.1 General

WARNING!
If any item is removed from the wall stand, e.g. the detector or lateral armrest, the wall
stand will become unbalanced.
When the brake is released, part of the wall stand will move upward and can cause
injury.
The operation should be performed by trained personnel.
Note!
The patient or operator is not allowed to hang or put any weight on the wallstand or patient
lateral armrest, they are only allowed to use it for support .
CAUTION!
If the voltage is not within the specified range, exposure problems may occur.
If the voltage is too low, the exposure may need to be repeated.
If the voltage is too high, the exposure may be impossible.
CAUTION!
Do not place the device where dust may cause malfunction to the power source.
CAUTION!
To avoid any injuries to patient, user or system, peripherals should always be placed
outside the working area.
CAUTION!
• Be careful not to load more than 25 kg on the lateral armrest.
• Patient must be supported by trained radiologist in using the lateral armrest.
• Do not use the lateral armrest when it is unlocked.
Operator´s Manual
2020-09-18
Operating the system
0180-095-418-EN Rev. 6.1
Page 103
Operating the Wall Stand
OMNERA®

Hide quick links: