Closed Table 0181 - Canon OMNERA 400T Operator's Manual

Manual-positioning digital radiographic system
Hide thumbs Also See for OMNERA 400T:
Table of Contents

Advertisement

8.1.6 Closed table 0181

8.1.6.1 Maximum patient load
Maximum patient load
8.1.6.2 Weight of parts
Table (with table top and vertical lift)
Table top
Vertical lift
8.1.6.3 Vertical lift
Lowest table top position (from the floor to
the table top surface)
Z stroke
Maximum travel speed
8.1.6.4 Table top
Dimensions
X-ray transparent area
Thickness
Length of stroke
X-direction from center position
(Longitudinal)
Length of stroke
Y-direction from center position (Lateral)
Aluminium equivalence
Aluminium equivalence cover detector
holder
8.1.6.5 Detector holder
Weight
Size (maximum)
Operator´s Manual
2020-09-18
Technical Specifications
295 kg
241 kg
47 kg
14 kg
540 +20/–10 mm
310 +40/–20 mm
25 mm/s (MRS ≥30 mm/s)
2400 mm X 800 mm
2350 mm X 580 mm
21 mm
±500 +20/–10 mm
±150 +20/–10 mm
≤0.9 mm
< 0.6 mm
Maximum 40 kg
D600 mm x W620 mm x H95 mm
0180-095-418-EN Rev. 6.1