Operator´s Manual
Technical Publication:
1000-095-071-EN Rev. 2.1

Advertisement

Table of Contents
loading

Summary of Contents for Canon OMNERA 500A

  • Page 1 Operator´s Manual Technical Publication: 1000-095-071-EN Rev. 2.1...
  • Page 2 All third-party brand names, product names, or logos appearing in this manual are the trademarks or registered trademarks of their respective owners. Canon® is a registered trademark of Canon, Inc. and may also be a registered trademark or a trademark in other countries.
  • Page 3 Date New release 2020-09 1.2 Identification Labels updated 2021–06 Updated 6 Cleaning and Disinfection to Added warning from 6 Cleaning and Disinfection to 2.2 Precautions, Safety Added 8 Complying Standards Added AEC Accuracy in table Output Parameter and Loading Factor in 9.3 Radiographic Specification Added 9.4 Collimator Added 9.5 X-ray Tube...
  • Page 5: Table Of Contents

    Direction of Movement................45 3.1.2 Sound Signal ..................45 3.1.3 System Display Overview ..............46 Automatic Collimator Control ................47 3.2.1 Automatic Collimator ................47 Display User Interface ..................51 3.3.1 Patient Information ................52 3.3.2 Position Information ................52 Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page i...
  • Page 6 Transport Interval Zone ................103 Manual Mode .................... 104 4.6.1 Activation of Manual Mode ..............104 Free Examination Procedures..............106 4.7.1 Free Mode ..................106 4.7.2 Auto Position Mode ................106 Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page ii...
  • Page 7 Collimator....................147 X-ray Tube ....................148 Environmental Requirements..............149 Ceiling Suspended X-Ray Tube Support............150 9.7.1 General ....................150 9.7.2 Configuration ..................150 9.7.3 Weight....................150 9.7.4 Electrical Characteristics ..............150 Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page iii...
  • Page 8 Grid....................158 12 Appendix A....................159 12.1 Glossary ....................159 13 Appendix B....................163 13.1 Monthly Checklist..................163 13.1.1 Overhead Tube Crane ................. 163 13.1.2 Remark ....................165 13.2 Annual Checks ..................166 Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page iv...
  • Page 9: Introduction

    • The original version of this manual is written in English. • Training is provided by or via Canon Medical Systems. Training material consists of the Operator’s Manual and the Installation and service manual.
  • Page 10: Text Emphasis

    Note! All texts labelled with “NOTE” contain additional information regarding the work step, and is provided for a better understanding or as a warning about unnecessary and avoidable difficulties. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 2...
  • Page 11: Identification Labels

    Introduction Identification Labels 1.2 Identification Labels The figure shows the location of the identification labels on the equipment. OMNERA 500A 1000 Fig. 1-1 Location of identification labels Label 13 and label 14 apply for US market only. Operator´s Manual 1000-095-071-EN Rev. 2.1...
  • Page 12 Introduction Identification Labels OMNERA 500A 1000 US only Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 4...
  • Page 13 Introduction Identification Labels 1000 US only Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 5...
  • Page 14: System Description

    Introduction System Description 1.3 System Description 1.3.1 General Omnera 500A includes: • Overhead tube crane (OTC) with x-ray tube and collimator • Table • Wallstand • System cabinet with a high voltage generator • Image Acquisition system • Flat panel detectors 1.3.2 Intended Use...
  • Page 15: System Overview

    Introduction System Description 1.3.4 System Overview Fig. 1-2 System Overview A Overhead tube crane D Wallstand B Table E System cabinet C Detector holder F Computer and monitor Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 7...
  • Page 16: Overhead Tube Crane, Overview

    1.3.5 Overhead Tube Crane, Overview Fig. 1-3 Overhead Tube Crane A Ceiling rail Y E X-ray tube B Traverse X F Maneuver handle C Ceiling wagon G Collimator D Column (Z) H Display Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 8...
  • Page 17 C Table top I Collimator hand control (option) D Table hand control (X/Y/Z, Ceiling tube pendulum movement) J Head end E Detector holder K Foot end F Brake release for detector holder Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 9...
  • Page 18: Wallstand Overview

    Fig. 1-5 Wallstand overview Wallstand with tilt option With non-tilt detector (standard) T= Tilt A. Lateral armrest E. Hand control (Collimator and movement B. Detector holder adjustments) C. Column D. Foot plate Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 10...
  • Page 19: Safety

    In particular such a separation device is required when a network connection is made. The requirements on the separation device is defined in IEC 60601–1, edition 3.1, clause 16. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 11...
  • Page 20: Precautions, Safety

    Use a lint-free cloth moistened with a moderate amount of liquid to avoid that cleaning liquids seep into the openings of the system, e.g., air openings, gaps between covers. • Do not restart the system if cleaning liquids have leaked in. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 12...
  • Page 21 Otherwise, the functions and performance of this device may be impaired. CAUTION! No objects shall be positioned within the working area. If necessary, they must be removable. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 13...
  • Page 22 Federal law restricts this device to be sold by or on the order of a physician. (US market only.) Note! Radio interference standard Federal Communications Commission (FCC) Part 15 Class B applies to this equipment. Note! The equipment may only be used as intended. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 14...
  • Page 23: Qualifications Of Personnel

    • are trained in the use of equipment and procedures of this type. Note! It is the responsibility of the owner to ensure that the technicians have the correct training and knowledge to perform service and maintenance. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 15...
  • Page 24: Service And Maintenance

    If national rules or regulations specify more frequent checks and/or maintenance, such regulations must be observed. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 16...
  • Page 25: Installation And Repair

    • components are not replaced by original spare parts in case of a malfunction. • the electrical installation of the room concerned does not meet the requirements or the corresponding national regulations. • the system is not used in accordance with the operating instructions. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 17...
  • Page 26: Safety And Warning Symbols

    Manufacturer Date of manufacture To indicate the emission or the imminent emission of X-radiation. Marking on the emergency stop button. Activation of the actuator interrupts all mechanical movements and prohibits exposures. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 18...
  • Page 27: Safety And Warning Labels On The Equipment

    25 kg = 55 lbs Max 25 kg 25 kg = 55 lbs Max 25 kg 300 kg = 661 lbs Max 300 kg Fig. 2-1 Locations of safety and warning labels Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 19...
  • Page 28: Applied Parts

    Safety Applied Parts 2.8 Applied Parts Applied parts are intended for the patient to touch. Fig. 2-2 Applied parts, System Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 20...
  • Page 29: Essential Performance And Basic Safety

    • The system can prevent a user initiated xray exposure to start if an error is detected that can affect essential performance or basic safety. • A termination of a user generated motorized movement. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 21...
  • Page 30: Emergency Stop

    When the emergency stop is activated, lift the overhead tube crane up and push or pull sideways to change the position of the tube/collimator manually. There are additional external emergency stops as option. Fig. 2-3 Emergency stop buttons Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 22...
  • Page 31: Radiation And X-Ray Tube

    Audio and visual communication must be possible between the operator and the patient when exposure is performed. Note! The X-ray beam should not be outside the boundaries of the detector holder. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 23...
  • Page 32: Radiation Protection

    • Use high kV and low mA to produce less scatter. • Collimate the exposure field to the area of interest. • Add collimator filter to reduce the scatter. • Compression of patient. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 24...
  • Page 33 100KV, 500mm 120KV, 1000mm 120KV, 1000mm, Lead apron 120KV, 500mm central Beam dose rate Fig. 2-4 Scattered radiation rate expressed in percent of central beam dose rate, with and without shielding Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 25...
  • Page 34 Fig. 2-5 shows a top view of the table and the zone of occupancy, where the arrows B show the direction of decreasing scatter radiation levels. Fig. 2-5 S = Significant zone of occupancy A Central beam B Decreasing Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 26...
  • Page 35 100KV, 1000mm, Lead apron 120KV, 1000mm, Lead apron 120KV, 500mm 120KV, 1000mm central Beam dose rate Fig. 2-6 Scattered radiation rate expressed in percent of central beam dose rate, with and without shielding Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 27...
  • Page 36 The remaining part of the X-ray beam after having passed the plane of the image reception area (detector and detector holder) can be significantly high. Never stand behind the wallstand during an exposure, see Fig. 2-7. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 28...
  • Page 37: Mechanical Safety

    Surrounding equipment is not subject of the collision warning. It is the operator’s duty, before any movements are activated, to ensure that any danger to the patient and/or third person is prevented. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 29...
  • Page 38: Overhead Tube Crane

    • Cover (2) and the column (Z) when the X- ray tube (3) is moving in beta (β) direction. 1. Column (Z) 2. Cover 3. X-ray tube 4. Column bottom plate Fig. 2-8 Overhead tube crane, mechanical safety Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 30...
  • Page 39: Cabinet

    Safety Mechanical Safety 2.12.3 Cabinet Cabinet, mechanical safety Max 25kg/ 55 lbs Max 25 kg/ 55 lbs Fig. 2-9 Placement of warning and safety label. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 31...
  • Page 40: Table

    Possible squeezing hazard areas and placement of warning labels: 002010.eps Fig. 2-10 Squeeze hazard 1. Table top (X/Y/Z) 5. Cover 2. Cover 6. Imaging unit rail 3. Column cover foot 7. Column (Z) 4. Imaging unit (X) 8. Footplate Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 32...
  • Page 41 The patients shall always have their extremities placed over the table top. Note! Do not lean against the floating table top. WARNING! Wheelchair patients shall always be placed outside the working area, when operating any motorized movement. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 33...
  • Page 42 To reduce the lateral forces on the table the operator should be placed on the opposite longitudinal side of the patient and the hospital bed. The operator (A) should drag the mattress with the patient from the hospital bed to the table. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 34...
  • Page 43 • Maximum load of at patient sitting on the table top: – 150 kg. Fig. 2-14 Maximum patient weight label The table frame is marked on the upper side with the maximum weight when positioning in outer positions, see Fig. 2-14. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 35...
  • Page 44 The working area comprises the table top including the stroke length of the table top in the X- and Y-direction. The measurements in the figure show the length of stroke in the X- and Y- direction. The dimensions have some tolerances and can differ from the manufacturer’s. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 36...
  • Page 45 600 mm Fig. 2-15 Table top stroke length 9 mm Fig. 2-16 Working area underneath table Fig. 2-17 Detector movement The detector movement is up to 850 mm, depending on detector type. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 37...
  • Page 46: Wallstand

    Maximum weight on the wallstand lateral armrest is 25 kg/ 55 lbs. 2.12.5.2 Working Area, Wallstand The working area of the wallstand is the area in front of the detector holder. Fig. 2-18 Working area, wallstand Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 38...
  • Page 47 Fig. 2-19 Possible squeeze hazards 1. Slide opening of detector wagon 2.12.5.4 Motorized Wallstand CAUTION! Patients shall be outside the working area when operating any motorized movement. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 39...
  • Page 48: Safety Functions

    To release the pressure, a movement in the opposite direction is allowed. When the pressure on the column has returned to normal, see Fig. 2-20, the warning message is removed and motorized movements are allowed again. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 40...
  • Page 49: Dead Man's Grip

    All buttons for motorized movements require constant activation. If the operator releases one of the buttons/controls, the System will immediately stop or engage the brakes (manual movements). The exposure hand control has the same functionality. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 41...
  • Page 50: Electromagnetic Compatibility (Emc)

    Do not place the system near MRI equipment or other equipment that generates a strong magnetic field. CAUTION! Mobile telephones and other radiating equipment can interfere with the function of the system and can therefore cause safety hazards. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 42...
  • Page 51 Guidance and manufacturer’s declaration - electromagnetic emissions. See 1000-095-072 Installation and Service Manual for tables including IEC 60601 test and compliance levels, and recommended separation distances between portable and mobile RF communications equipment and system. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 43...
  • Page 52 Safety Electromagnetic Compatibility (EMC) Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 44...
  • Page 53: User Interface

    • One sound signal means that the overhead tube crane is in position and ready for exposure. • Two sound signals, in rapid succession, indicates a fault and the display shows an error message, for example after a collision. The error message shows the corrective action. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 45...
  • Page 54: System Display Overview

    9. Light indication, see 4. X direction 3.3.12 Light Indication 5. Emergency brake (rear side) 10. Display user interface, see 6. Release X-Y-direction and rotation 3.3 Display User Interface (alpha/beta) (rear side) Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 46...
  • Page 55: Automatic Collimator Control

    Imaging system. This can be useful if the collimator area is intended to be kept for the next examination. 2. Collimator light/x-ray field height x width 3. Collimator filter selection Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 47...
  • Page 56 • Changes to Manual mode. Changes: Size: 30 cm x 20 cm, filter 2. • Changes back to Automatic mode: Size 30 cm x 10 cm, filter 1. When a new patient is selected Automatic mode is automatically activated. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 48...
  • Page 57 In Free mode/Auto position mode the position of the detector is unknown for the System. The indicated size of the light field is correct at the shown SID. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 49...
  • Page 58 F. Button for closing the format width collimation. 3.2.1.7 Hand Control, Wallstand – Collimator Adjustment A. Collimator light on/off B. Adjustment height collimation C. Adjustment width collimation Fig. 3-8 Hand control Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 50...
  • Page 59: Display User Interface

    9. Selection of active AEC field (AEC mode only) 10. Patient size 11. Collimator centering 12. Manual or Servo mode 13. Density See the following pages for detailed description of the functions. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 51...
  • Page 60: Patient Information

    Fig. 3-21. Fig. 3-12 Position information A Alpha angle (°) B Beta angle (°) C Source Image Distance (SID), or Height to floor (H) in Free or Auto Position Mode (cm/inch) Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 52...
  • Page 61: Adjustment Of Generator Parameters (Kv, Ma, Ms, Mas, Density)

    • Select to increase or decrease the value. Fig. 3-13 Adjustment of generator parameters Note! The Operator/User is always responsible for checking and validating the exposure parameters in the Image system before performing exposure. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 53...
  • Page 62: Settings

    • USER SETTINGS • SERVICE The USER SETTINGS menu has the following tabs: • DISPLAY • SETTINGS • THEMES The SERVICE menu has the following tabs: • LOG • SETTINGS • DISPLAY Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 54...
  • Page 63 Patient info is shown when pushing the black field with the Abdomen Suspine Fig. 3-19 Always on not selected. The Patient info closes automatically or by pushing the black field with the again. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 55...
  • Page 64 – Auto Position # By selecting Auto Position # the auto position # is shown in the OTC display. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 56...
  • Page 65 Jane Doe ID 987-65-4320 Hand PA Density post mAs Fig. 3-23 Preview image enlarged Use the arrows appearing in the image to pan in the image. Fig. 3-24 Zooming In/Out Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 57...
  • Page 66 Brightness Logotype Fig. 3-25 Menu USER SETTINGS – tab SETTINGS • Use LCD to adjust the display brightness. • Use Logotype to set if the logo shall be shown or not. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 58...
  • Page 67 All, or just warnings and errors, Warnings & Errors. By selecting Warning, Error, or Information in the right column, more information concerning the issue is shown. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 59...
  • Page 68 Fig. 3-28 Delete log file Service – Settings The Service menu shows system set up and system software versions. A code is required for access to Settings. Fig. 3-29 Menu SERVICE – tab SETTINGS Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 60...
  • Page 69: System Mode

    Automatic positions are a number of positions that can be programmed and saved into the System. The stand moves upon activation of the servo button to the programmed position chosen from the imaging unit. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 61...
  • Page 70 Exposure Validation Exposure is allowed if the stand is not moving, operating properly (not in an error state) and the servo button is activated. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 62...
  • Page 71 Exposure is allowed if the stand is not moving, operating properly (not in an error state), the X-ray tube is aimed to the center of the detector and the servo button is activated. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 63...
  • Page 72 Exposure is allowed if the stand is not moving, operating properly (not in an error state), the X-ray tube is aimed to the center of the detector and the servo button is activated. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 64...
  • Page 73 Stitching is possible at both table and wallstand. Fig. 3-31 Composite Image A Partial Image B Partial Image Height C Overlap D Composite Image Height Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 65...
  • Page 74: Hospital Method Book

    The Hospital method book is reached by a activating the Hospital method book button for 1 second. The Hospital method book is selectable when a method book has been loaded to the display (performed by a service technician). Fig. 3-32 Hospital method book button Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 66...
  • Page 75: Selection Of Technique Mode

    To avoid unnecessary radiation, make sure that the AEC back-up values are properly defined. Density AEC Backup Fig. 3-34 mAs selection button grayed out For more detailed information about the different technique modes, please see Operator’s Manual for Canon single console CXDI NE. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 67...
  • Page 76 The AEC fields are activated by selecting them in the pop-up window 2 to the right. All activated AEC fields are shown at 1. AEC fields are deactivated by selecting them again in the pop-up window 2. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 68...
  • Page 77: Patient Size

    The generator parameters and the collimator settings (field size and filter) change to the defined values for the new Patient size. If no generator parameters or collimator settings are defined for the new Patient size (defined in APR), the current values are kept. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 69...
  • Page 78: Collimator Centering

    Bottom centering. Select the desired collimator centering. The pop-up window closes automatically short after the selection and the light field is accordingly adjusted. A. Top B. Center C. Bottom Fig. 3-39 Collimator centering selection Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 70...
  • Page 79: Servo State Mode

    For further information about Manual mode, Fig. 3-40 Servo state mode see corresponding section. A Automatic mode B Manual mode Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 71...
  • Page 80: Grid Status

    When a grid is not needed for the examination "Removed" is shown with black letters in the same location, see Fig. Fig. 3-42. When a correction of grid status is needed this is indicated with red text in the Canon NE interface, see detailed description in Table 3-1.
  • Page 81 115cm_10:1_52 in the image system user image system user lp/cm interface). Change to the interface guides to correct grid exchange the grid inserted to the requested grid according to the protocol. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 73...
  • Page 82: Light Indication

    • Yellow flashing — Action needed by the user or system is moving • Green flashing — System is ready for exposure • Green fixed — Preparation (before exposure) • Yellow fixed — Exposure Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 74...
  • Page 83: Exposure

    The deviation index, DI, gives an indication of the dose level used for capturing the image. The DI value compares the current standard EXI with the target EXI. The target EXI is defined by the user. See Imaging system for further description. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 75...
  • Page 84: Image System

    User Interface Image System 3.5 Image System For information about Image system functions, see the Image System Manual. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 76...
  • Page 85: Remote Control (Option)

    Servo button The servo button is yellow with a little peg, making it easy to recognize the button. When activating the yellow servo button, the overhead tube crane moves to auto-position. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 77...
  • Page 86 The remote control uses 2 LR03, 1.5V, AAA batteries. To change batteries, loosen the 3 screws and open the slot at the back of the remote control. Note! The batteries shall be recycled. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 78...
  • Page 87: Servo Button

    If a DAP meter is included in the System, the Dose Area Product is presented in the Imaging system. Checks and settings can be done by the service software, see ‘Installation and Service Manual’, Chapter 5, Installation. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 79...
  • Page 88: Directions Of Movement

    B. Press the button to release the brakes on the table top, Y and X. On activation, the table top can be moved manually. When the release button is activated, the collimator light is lit. C. Press button to move the table top upward. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 80...
  • Page 89: Foot Control, Wireless (Option)

    1 week of constant use, or 168 hours. It then changes to 2 flashes per second when the capacity has been reduced to approximately 2 days, or 48 hours. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 81...
  • Page 90: Xy Foot Control, Strip Type (Option)

    Press and hold the foot control strip type (1) to release the brakes (X, Y) on the table top (2). When activated, the table top can be moved manually (floating table top). Fig. 3-51 Foot control strip type 1. XY foot control strip type (option) 2. Table top (X/Y) Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 82...
  • Page 91 E. Drive motorized image receptor holder to right. F. Move the overhead tube crane to the left in Pendulum mode. G. Move the overhead tube crane to the right in Pendulum mode. H. Not used. I. Not used. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 83...
  • Page 92: Moving Table Top

    The Table may, as an option, be equipped with a vertical travel safety system that protects the table top. If a collision is detected and the force exceeds 20 kg, all movements are stopped. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 84...
  • Page 93: Attach/Remove Accessories

    3. Click to attach at A. Fig. 3-55 Attach accessories • To remove: 1. Press on the accessory at B. 2. Rotate it upward. 3. Remove the accessory. Fig. 3-56 Remove accessories Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 85...
  • Page 94: Motorized Imaging Unit Movement

    The correct position is indicated by the lightning of the green detector holder button. Note! It is the user’s responsibility to verify that the detector is in position at exposure. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 86...
  • Page 95: Wallstand Control Elements

    Adjustment of the automatic collimator, vertical movement of the detector, rotation of the detector and activation of pendulum mode. The wallstand detector is tilted by activation of the hand control. The detector can be tilted -20 to + 90° . Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 87...
  • Page 96 E. Break release for manual movement of detector F. Pendulum mode wallstand G. Detector up/down, Motorized H. Detector tilt and overhead tube crane tracking, -20 to 90° Fig. 3-58 Hand control Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 88...
  • Page 97 Note! Armrest has to be removed to allow tilt movement. 3.9.1.2 Light Indication (A) The selected Workstation is indicated in the imaging system and with a green light on the wallstand. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 89...
  • Page 98 A. Press the pedal to move the detector downward. B. Press the pedal to release the brakes. When activated, the detector can be moved manually. C. Press the pedal to move the detector upward. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 90...
  • Page 99: Operating The System

    When the system starts up, light indications and displays are lit. • Perform the following procedure when starting up the X-ray system: 1. Press the power [ON] button (A) on the mini console. Fig. 4-1 Power on button – mini console Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 91...
  • Page 100 5. Log in on the computer. 6. Type user name and password, press Log in. Fig. 4-4 7. Confirm that the image system has started normally by checking the status icons. Fig. 4-5 Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 92...
  • Page 101: Turn Off The System

    1. Move the overhead tube crane to a parking position, for example over the table. 2. Turn off the image system Select Fig. 4-6 Fig. 4-7 Shut down Confirmation Dialog Box Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 93...
  • Page 102 4. Press the power [OFF] button (B) on the mini console. It is possible to turn off the power to the X-ray system while the power to the image system is still on. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 94...
  • Page 103: Perform Examination

    Operating the System Perform Examination 4.3 Perform Examination 4.3.1 Select Patient 1. Select Exam and Worklist. 2. Sort the list in Search For Study List and select patient. Fig. 4-9 Select Worklist Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 95...
  • Page 104: Start Examination

    Perform Examination 4.3.2 Start Examination 1. Select Start Exam. Predefined protocols are activated automatically. Patient data can also be added manually, see Canon Operation Manual. Fig. 4-10 The indication light is lit on the selected workstation. 4001 Fig. 4-11 Lights...
  • Page 105: Position The System

    • 3.8 Table Control elements, Page 80 • 3.9 Wallstand Control Elements, Page 87 The light field should be reduced to the examination area. Adjust the collimator according to: • 3.2 Automatic Collimator Control, Page 47 Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 97...
  • Page 106: Exposure

    It is the responsibility of the user to ensure that the X-ray field is within the active detector area when exposing. CAUTION! Materials located in the X-ray beam may cause adverse image effects. Note! The operator is responsible for verifying the exposure parameters before exposure. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 98...
  • Page 107: Review Image

    • If the image is correct, select End Exam or continue the examination if more images are planned. • If the image is not correct, select Retake. • Type reject reason and retake image. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 99...
  • Page 108: Change Work Space

    C. Patient posture (detector setup) D. Stand E. Table F. Cassette G. Universal H. Available workspaces I. Incomplete detector connection or turned off power box J. Workspace name K. Close Fig. 4-16 Detector or workspace Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 100...
  • Page 109: Basic Exposure Error Handling

    Place the detector in the Wallstand examination Is the detector in the wallstand detector holder, wallstand detector holder? make sure to connect the connector correctly. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 101...
  • Page 110: Movement Short-Cut Zones

    The wallstand short-cut zone is defined in front of the wallstand. Fig. 4-17 Wallstand short-cut zone 4.4.2 Table Short-Cut Zone The Table short-cut zone is defined above the table top. Fig. 4-18 Table short-cut zone Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 102...
  • Page 111: Transport Interval Zone

    Inside the transport interval zone the overhead tube crane moves horizontally until it reaches the intended X, Y-position. Then the overhead tube crane moves vertically to the intended Z-position. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 103...
  • Page 112: Manual Mode

    Manual mode is selected. In the same way, the wallstand is shown without connection to the overhead tube crane in Manual mode, see Fig. 4-21 Fig. 4-21 Wallstand and Table shown without connection to the overhead tube crane in Manual mode Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 104...
  • Page 113 SID value is not able to be calculated is some positions. No SID value is shown if the System is not in an active servo mode when Manual mode is activated. Tracking or other mode movements are not possible when Manual mode is active. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 105...
  • Page 114: Free Examination Procedures

    The Auto position mode is designed for emergency examinations or examination with a mobile detector. 4.7.2.2 Pre-defined Auto Positions When the auto position mode is selected, the overhead tube crane is go to a predefined position in the room. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 106...
  • Page 115 • The overhead tube crane moves automatically to its programmed position. • The servo mode button light indication is switched off. All movements are available. 4.7.2.4 Exposure Exposure is possible when the overhead tube crane is not moving. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 107...
  • Page 116: X-Ray Table Examination Procedures

    Exposure is possible when the overhead tube crane is not moving, and the servo mode indication light is fixed. Note! In “Table Flexible mode” examinations exposure is possible outside the imaging unit. The imaging unit does not follow the overhead tube crane. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 108...
  • Page 117: Film Tracking Mode

    SID. 4.8.2.3 Exposure Exposure is possible when the overhead tube crane is not moving, and the servo mode indication light is fixed, and the x-ray beam covers the image unit Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 109...
  • Page 118: Pendulum Mode

    Fig. 3-53. The tube move in the desired direction and the imaging unit move to stay aligned with the tube. 4.8.3.3 Exposure Exposure is possible when the overhead tube crane is not moving, and the servo button is activated. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 110...
  • Page 119: Stitching Table Mode (Option)

    Press the button (F1) to set the right limit, see Fig. 4-28. • The button turns green to indicate that the limit is set. Fig. 4-28 Set left/right limit Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 111...
  • Page 120 • After exposure, the system moves to the next correct position and the second image is captured. This is repeated until all images for the composite image has been captured. • The system beeps when the sequence is finished. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 112...
  • Page 121: Wallstand Examination Procedures

    • Beta and Sideways blocked — It is only possible to move the overhead tube crane up and down. Beta and Sideways blocked – No sideways movements allowed. Blocked non- FFD directional (X/Y) movement, movement is turn servo off and block exposure. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 113...
  • Page 122 • The overhead tube crane turns on the collimator light when the correct height is reached. Adjust the position. • The mode stays activated. • The SID value is changed to the new distance (if selected distance toward the wallstand is changed). Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 114...
  • Page 123: Nowait Configuration

    • The SID value is changed to the new distance (if selected distance toward the wallstand is changed). 4.9.2.1 Exposure Exposure is possible when the overhead tube crane stands still, and the servo mode indication light is fixed. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 115...
  • Page 124: Stitching Wallstand Mode (Option)

    Z direction. 6. Rotate the x-ray tube in order for the lower edge of the collimator light field to indicate the lower limit for the composite image. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 116...
  • Page 125: Aec Technique Setup

    • This is repeated until all images for the composite image has been captured. • The System beeps when the sequence is finished. Release the button. 4.9.4 AEC Technique Setup For information about the AEC Technique setup, see the Generator Manual. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 117...
  • Page 126: Detector, Wallstand

    • Depending on left or right operated wallstand, the location of the detector tray and position of button and latches is different. Note! If the System includes more than one detector, assure that the active detector is used. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 118...
  • Page 127: 14X17 Inches Detector, Wallstand

    2. Insert the detector into the detector holder, as shown below and set it by pushing it in, until it clicks. Fig. 4-37 Insert detector CAUTION! • Confirm that the latch is going up firmly, as shown below. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 119...
  • Page 128 Operating the System Detector, Wallstand 3. While pressing and holding the button of the detector tray, return it back to the inside of the detector holder. Fig. 4-38 Return detector tray Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 120...
  • Page 129 In the landscape position, there is a choice between having the detector rotated through the detector center or through the detector top. 1. Center position landscape 2. Top position landscape 3. Portrait Fig. 4-40 Rotating detector Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 121...
  • Page 130 Note! • Depending on left or right operated wallstand, the location of the detector tray and position of button and latches is different. Fig. 4-41 Unlocking detector latch Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 122...
  • Page 131: 17X17 Inches Detector, Wallstand

    Fig. 4-42 Pull out detector tray 2. Insert the detector into the detector tray as shown below and set it by pushing the detector, holding down the latch. Fig. 4-43 Insert detector Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 123...
  • Page 132 It is important that the hold-backs lock outside the detector end. Failure to position the detector in the proper position leads to incorrect images. Note! If the detector or the detector holder is not properly inserted, a warning symbol is shown at the display. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 124...
  • Page 133: Portable Detector, Table

    If the System includes more than one detector, assure that the active detector is used. 4.11.1 Set the detector 1. Pull out the detector tray. Fig. 4-46 Releasing the detector tray Fig. 4-47 Pulling out the detector tray Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 125...
  • Page 134 4. Press and hold the button of the tray and push it in. Fig. 4-50 Reinserting the detector tray To check if the detector is in the right position, see the “STATE” ; Detector position/ present/undefined at the display. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 126...
  • Page 135: Change Between Portrait And Landscape

    4.11.2 Change Between Portrait and Landscape 1. Rotate the detector 90°. 2. Hold as shown below and turn the detector, clockwise or anti-clockwise. Fig. 4-51 Detector change between portrait and landscape Fig. 4-52 Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 127...
  • Page 136: Remove Detector

    1. Withdraw the detector holder and rotate the detector, if needed, to remove the detector. Fig. 4-53 2. Remove the detector by pulling it toward you according to Fig. 4-54. Fig. 4-54 Detector removal Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 128...
  • Page 137: Exchange Grid

    Insert the grid with the tube side facing up, toward the X-ray source. The tube side of the grid has the specification label and the grid centre line identification. 3. Push in the grid, until it clicks. Fig. 4-56 Push in the grid Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 129...
  • Page 138 Operating the System Portable detector, table Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 130...
  • Page 139: Error Handling

    Reason: 0xXXXX Component: 0xXXXX Extra: 0xXXXX 1 / 2 ERROR Fig. 5-1 Error pop-up window The Error pop-up window disappears when the user pushes the close button. Fig. 5-2 Close button Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 131...
  • Page 140 Fig. 5-6 Close button 5.1.1.3 Log The Log file is part of the Setting menu and reached by pressing the gear or the Error/ Warning messenger bars. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 132...
  • Page 141: Diagnostic

    Error Handling Diagnostic 5.2 Diagnostic For System messages, error messages and error handling, see the System Installation and Service Manual. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 133...
  • Page 142 Error Handling Diagnostic Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 134...
  • Page 143: Cleaning And Disinfection

    For cleaning and disinfection of the following parts, the system can stay switched on: • Lateral armrest • Patient grips • Chin rest • Front cover of Bucky unit • Tabletop • Maneuver Handle and Display, see 6.4. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 135...
  • Page 144: Cleaning

    • Remove all remaining cleaning residues and dry with a soft cloth. • Keep the ventilation slots of all components unobstructed. • Regularly clean the dust off all rails and joints etc. See also separate instruction for 6.4 Maneuver Handle and Display. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 136...
  • Page 145: Disinfection

    • Wipe the surface with a lint-free cloth moistened with a disinfectant. • Do not spray any disinfectants directly on the surface. • Obey the instructions of the manufacturer of the disinfectant. See also separate instruction for 6.4 Maneuver Handle and Display. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 137...
  • Page 146: Maneuver Handle And Display

    • Wipe the maneuver handle and display using a moderately moist cloth with water or alcohol-only cleaning agents only. • Do not spray directly on the maneuver handle or the display. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 138...
  • Page 147: Function And Safety Checks

    MS: Allows the operator to adjust the AEC backup ms, not to exceed preset AEC backup mAs/ms values or system limits. The ms value will be displayed in the time window of the LED display during AEC operation. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 139...
  • Page 148: Monthly Checks

    9. Move the table top longitudinal and check that the mechanical end stops are not loose. 7.2.4 Wallstand 10. Move the wallstand up and down in Z direction and make sure it runs smoothly and sounds OK. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 140...
  • Page 149: Complying Standards

    – Medical electrical equipment: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis. IEC 60601-2-54:2009+AMD1:2015+AMD2:2018 – Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 141...
  • Page 150 Complying Standards Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 142...
  • Page 151: Technical Specification

    480VAC 3Phase, +/-10%, 50/60Hz Long-time (positioning) 2 A 50/60 Hz. Momentary (exposure):150 A, 50/60 Hz (Short term peak value), (recommended fuse 63 A, thermal breaker, B curve.) Class 1 Heat dissipation 1713 BTU/H Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 143...
  • Page 152: Power Line Requirements

    50 kW 80 A 0.24 Ω 65 kVa 480 VAC, 3p 65 kW 105 A 0.19 Ω 85 kVa 480 VAC, 3p 80 kW 130A 0.15 Ω 105 kVa 480 VAC, 3p Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 144...
  • Page 153: Radiographic Specification

    10 to 800 mA (65 kW) 10 to 1000 mA (80 kW Variable in 0.1 mAs steps via protocol: mA steps: Variable according to ISO 497 Series R'20 via console Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 145...
  • Page 154 The AEC ms range is 15 ms to an installer-programmable maximum not to exceed 600 mAs. All models Coefficient of variation of AEC Accuracy measured air kerma ≤ 0.05 Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 146...
  • Page 155: Collimator

    Min: 0.3 x 0.3 cm Square X-ray field selection Max: 48 x 48 cm (± 1 % at 1 m SID ) (to reach 43 x 43 cm at 90 cm SID ) Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 147...
  • Page 156: X-Ray Tube

    Inherent filtration 0.7 mm Al/75KV Added filtration 0.8 mm Al 1.5 mm Al (0.7+0.8) Total filtration For more detailed x-ray tube technical specifications, see the provided tube insert and housing datasheets. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 148...
  • Page 157: Environmental Requirements

    10-90%, non-condensing Operating humidity (relative) 30-75% RH, non-condensing Maximum transport and storage altitude 3000 m Maximum operating altitude 3000 m Maximum air pressure 700–1060 hPa 55dB or less (except single noise) Noise Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 149...
  • Page 158: Ceiling Suspended X-Ray Tube Support

    ON / 4 min. OFF The equipment is not suitable for use in the Use of anaesthetic mixtures presence of flammable anaesthetic mixtures with air or with oxygen or with nitrous oxide. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 150...
  • Page 159: Speed

    60 mm/s X movement 250 mm/s 500 mm/s Y movement 250 mm/s 500 mm/s α movement 16°/s β movement 16°/s Image receptor holder 166 mm/s 350 mm/s movement (with 50 kg mass) Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 151...
  • Page 160: Cabinet

    Technical specification Cabinet 9.8 Cabinet Dimensions (L x W x H) mm 750 x 600 x 1125 mm Weight Max 134 kg Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 152...
  • Page 161: Table

    9.9.4 Electrical Characteristics Maximum power without external 500 W electronics 9.9.5 Attenuation Equivalent Table top < 0.9 mm AL at 3.7 mm HVL Detector holder ≤ 0.6 mm AL at 3.7mm HVL Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 153...
  • Page 162: Wallstand

    Rotation range detector holder wagon (Only the tiltable detector holder wagon). -20° - 90° Attenuation Equivalent Detector holder ≤ 0.6 mm AL at 3.7mm HVL Weight 200 kg ±10 Maximum speed Z movement 200 mm/s Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 154...
  • Page 163: Waste Disposal

    For disposal of other components, refer to corresponding documentation. Follow the rules and regulations of your relevant authorities in the disposal of this system, accessories, options, consumables, media and their packing materials. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 155...
  • Page 164 Waste Disposal Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 156...
  • Page 165: Accessories

    0072-099-011 Lateral cassette holder X, Y, Z Foot control 0072-099-004 X, Y Foot control strip type 0055-099-025 Compression belt, cost effective 0072-099-028 Compression belt, high-end 0072-099-029 Form pad, small-rectangle 0080-099-051 Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 157...
  • Page 166: Wallstand

    Grid 52 lp/cm, 10:1 Ratio, F140, Al type 0180-099-076 Grid 52 lp/cm, 10:1 Ratio, F180, Al type 0180-099-061 Grid 52 lp/cm, 10:1 Ratio, F115, Carbon 0180-099-082 cover Grid 52 lp/cm, 10:1 Ratio, F180, Carbon 0180-099-083 cover Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 158...
  • Page 167: Glossary

    Electrical component that leads voltage and current in one direction. Diode See “Supplier”. Dealer Image receptor for X-ray that does not require a cassette. The Detector reception and transfer of an image is digital. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 159...
  • Page 168 A control and supervision unit, consists of printed circuit board and Node node specific software. Optic Density. O.D. Extra facilities that demand updating of the System software and Options hardware before use. Options demand installation of an authorized service technician. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 160...
  • Page 169 The size of the Table top including X- and Y-stroke. The System moves in the X-direction. X-movement The System moves in the Y-direction. Y-movement Z-node The Z-node controls the Z-movement. The System moves in the Z-direction. Z-movement Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 161...
  • Page 170 Appendix A Glossary Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 162...
  • Page 171: Monthly Checklist

    5. Check the hoses for damage. 6. Check all outer cabling for damage. 7. Clean all outer surfaces, except for the lubricated column segments. See Chapter “Cleaning” at the Instruction for use. Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 163...
  • Page 172 9. Move the Table top in longitudinal direction and check that the mechanical end stops are not loose. 13.1.1.4 Wallstand 10. Move the wallstand up and down in Z direction and make sure it runs smoothly and sounds OK. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 164...
  • Page 173: Remark

    Appendix B Monthly Checklist 13.1.2 Remark Remark Action Int Note Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 165...
  • Page 174: Annual Checks

    Appendix B Annual Checks 13.2 Annual Checks Refer to Service and Installation Manual. Operator´s Manual OMNERA 500A 1000-095-071-EN Rev. 2.1 Page 166...
  • Page 175 Appendix B Annual Checks Operator´s Manual 1000-095-071-EN Rev. 2.1 OMNERA 500A Page 167...