5.8

Post-installation checks

1)
Check sheet
Checkpoint
Checking the imaging unit
Checking the date and time
Checking the software version
Identifying and registering the
imaging units
Inputting the control PC serial
number
Checking the operation unit
Checking the exposure
condition table
Checking the annotation
Network connections
Preparations prior to exposure
Checking image transfer to
printers and storages
Checking image transfer to
external memory device
Checking the image quality
Deleting unnecessary data
(there must be no unnecessary
data such as the images used
for testing)
Checking the window displays
(no unnecessary windows must
appear; the same applies after
rebooting)
2. Installation Manual
Checkpoint details
Align the unit with the X-ray tube
Check that the unit does not interfere with the
cables.
Set the date.
Set the time.
Set the time zone.
Check that the CXDI application, firmware and
PLD code versions all match.
Register the serial numbers of the sensors
Input the serial number of the control PC to be used.
Set the contrast.
Set the brightness.
Set the gamma correction.
kV
mA
msec or mAs
Check that the setting have been made in
accordance with the customer's request.
IP address
Subnet mask
Default gateway
Perform calibration.
Perform self-test.
Check that the setting have been made in
accordance with the customer's request.
Check that the image is transferred properly.
Use SMPTE pattern to check the density on a linear
chart.
Check that there is no artifacts, shading, etc.
dtque
dtstore
dttmp
old
Windows XP trash box
Operate from the Windows XP desktop.
Taskbar
- 87 -
Check