User Interfaces
Automatic collimator (option)

3.15.3 Display and control elements

3.15.3.1 Display automatic collimator
The figure shows the functions of the automatic collimator.
1. Adjusting knob for format height collimation (turning to the left closes the collimator, turn-
ing to the right opens the collimator).
2. Adjusting knob for format width collimation (turning to the left closes the collimator, turn-
ing to the right opens the collimator).
3. Button for switching the light and the laser line on/off. The light and the laser line is auto-
matically switched off via a time switch.
4. Measuring-tape grip for SID measurement. The measuring tape has both a cm and an
inch graduation.
5. Detent lever for ±45° rotation of the collimator around the central beam axis. The collima-
tor stops in the 0° position.
6. Button for changing between automatic and manual mode. A long activation of the M but-
ton will set the light field to the detector size if tracking WS or Table is active. If no tracking
is active, a long activation will set the light field to maximum size and automatic mode.
7. Two accessory rails.
8. Function display will indicate manual or automatic mode (ACSS) of the collimator. The
display will also show the select pre-filtration, the size of the light field and the pre-pro-
grammed SID value.
9. Button for manually changing the SID. The new SID value will then be used for calculating
the field size instead of the pre-programmed value, steps: 100, 115, 150, 180, 200.
10. Button for selecting collimator pre-filtration.
11. Control for covering the laser line.
OMNERA®
Manual
0.2mmCu
35.0cm x 35.0cm
11
Fig. 3-34 Display and control elements
0180-095-418-EN Rev. 6.1
Page 70
100cm
Operator´s Manual
2020-09-18