Checking Startup and Shutdown of Unit

Turn ON the machine to check that it starts up and shuts down properly.
This is the final step in the process of checking startup and shutdown of the unit.
operation
erase
"operation" switch
2-25

2.8 Checking Startup and Shutdown of Unit

1
Turn ON the facility power supply.
2
Turn ON the power breaker of the unit.
3
Press the "operation" switch.
4
Check the following:
The message display window shows the message
"INITIALIZING" and a progress indicator bar.
The bar extends to the right during the initialization
process.
When the bar reaches the right end, "NETWORK
ERROR [24000]" appears in the upper part of the
message display window.
The factory-set IP address of the machine is
displayed in the lower part of the message display
window.
5
Press the "operation" switch for at least three
seconds.
The backlight of the message display window goes off.
6
Turn OFF the power breaker.