3.3.4 Checking Safety Interlock Operation

The unit is equipped with the following safety interlock. Check that the interlock mechanism operates
correctly.
Position
Second front cover
Check that opening the second front cover during initialization stops the driving operation.
1
Open the second front cover when the unit is in READY status.
2
Close the second front cover.
The initialization process starts.
3
Open the second front cover after the driving operation starts.
The driving operation stops when the interlock mechanism works properly.
4
Close the second front cover.
The initialization process starts again. When it finishes, "READY" appears in the message display
window.
This is the final step in the process of checking the safety interlock operation.

3.3.5 Backing Up Setup Data

Back up the unit setup data in the REGIUS console.
The setup data for the unit is stored on the CF card mounted on the CIU circuit board.
Note
For information about the procedure for backing up the data, refer to the REGIUS console Installation/
Service Manual.
This is the final step in the process of backing up the setup data.
3.3 Checking Operations and Backing Up Setup Data
Condition to be Detected
Opened door while the
power supply is ON
3-7
Machine Status Caused by Interlock
Turns OFF the driving power supply
Shuts down the laser power supply
Interlock switch