3.3

Checking Operations and Backing Up Setup Data

Check whether the unit operates in the system to read images successfully.
It the machine operates successfully, back up the unit setup data in the REGIUS console.

3.3.1 Checking Images

Use a solid image to check whether it has been read successfully. At the same time, check that the unit
operates correctly.
For information about the procedure for checking images, refer to the REGIUS console Installation/
Service Manual.
This is the final step in the process of checking images.

3.3.2 Checking Coordinative Operations

Check that the image is read on all REGIUS consoles connected to the unit.
For information about the operation procedure, refer to the REGIUS console Operation Manual.
This is the final step in the process of checking coordinative operations.
Proceed to

"3.3.3 Setting Sound Volume"

When you do not need to change the sound volume, proceed to
Operation"
to check that the interlock mechanism of the unit operates correctly.

3.3.3 Setting Sound Volume

The speaker sound volume is set at low level at the time of shipment.
To increase the sound volume, change the "Volume" setting on the "Settings" panel of the "Unit Test"
screen. (Three setting values are available: "1 (Low)" (Default), "2 (Normal)", and "3 (High)".)
For information bout the "Settings" panel and the procedure for changing the setting, refer to "Service Tool
Screen (Unit Test)" in the unit Service Manual.
This is the final step in the process of setting the sound volume.

3.3 Checking Operations and Backing Up Setup Data

to change the sound volume setting for the speaker of the unit.
"3.3.4 Checking Safety Interlock