6.4 Combi MODULE

6.4.1 Replacing the sieve in the oxygen inlet

See "6.2.1 Replacing the sieve in the oxygen inlet" on page 13.

6.4.2 Replacing the O-ring in the oxygen outlet

See "6.2.2 Replacing the O-ring in the oxygen outlet" on page 14.

6.4.3 Dismantling the housing

See "6.2.3 Dismantling the housing" on page 14, with
Re: Combi MODULE has two toggle switches:

6.4.4 Assembling the housing

See "6.2.4 Assembling the housing" on page 16, with
1. Re: step
:
4.
Combi MODULE only requires one blind
plug
.
35
2. Re: step
:
11.
As Combi MODULE has two toggle switches,
please use a guide pipe over both toggle swit-
ches when mounting the upper housing
section
.
15
:
one exception
:
two exceptions
35
15
20
29
Combi MODULE

Advertisement