4.2
Routine Periodic Maintenance
Maintenance for the Model Mini-ICS Oxygen analyzer is very simple. Apart from the normal
maintenance for any instrument, such as cleaning the chassis, wiping the display, and replacing
the sensor, the Model Mini-ICS does not require any major servicing. Periodic calibration of the
sensor on known gas should be performed on a regular basis (see section 4.2). The chart below
should serve as a general guide for maintenance personnel.
TASK
CALIBRATE SENSOR
CLEAN THE ANALYZER
CHASSIS AND
DISPLAY PANEL WITH
SOFT CLOTH. MAKE
SURE THE
VENTILATION PORTS
ARE CLEAR.
CONFIGURE ALARMS
CHECK THE ANALOG
OUTPUTS AND RS-232
OUTPUT AGAINST
DISPLAY
INSTALL NEW OXYGEN
SENSOR
Figure 13 – Maintenance schedule
Manual Part Number:
5-06-4900-53-0
RECOMMENDED FREQUENCY
COMMISSIONIN
MONTHL
G
Revision Level: C
ECO 044952
ANNUALL
Y
Y
Revision Date:
May 08, 2018
Troubleshooting
AS REQUIRED
• GP/IT Sensors
9-12 Mos.
• CAG250
Sensors 2-3
Yrs.
Page
4-19