4.1.2.10.6 Calibrate 0 To 1 Volt Output
This menu calibrates the 0-1V Analog Voltage Output. A digital multi-meter set to measure volts
must be attached to TB3-pin5 and TB3-pin6. Jumper RA must be open and RB must be shorted
(section 4.1.3). Navigate from Analog Output Setup menu to the Analog Output Calibration menu
by typing "3" on the RS-232 terminal. Then, type "4" to access the Calibrate 0 To 1 Volt Output
menu.
4.1.2.10.7 Calibrate 0 To 20 Milliamp Output
This menu calibrates the 0-20mA Analog Current Output. A digital multi-meter set to measure mA
must be attached to TB3-pin7 and TB3-pin8. Navigate from Analog Output Setup menu to the
Analog Output Calibration menu by typing "3" on the RS-232 terminal. Then, type "5" to access
the Calibrate 0 To 20mA Output menu.
4.1.2.10.8 Oxygen Concentration at Voltage Manual Zero
This menu sets the oxygen concentration at which the analog voltage output is at minimum scale.
This setting may be set anywhere from 0.0 to 100% oxygen. Navigate from Analog Output Setup
menu to the Analog Output Calibration menu by typing "3" on the RS-232 terminal. Then, type "6"
to access the O2 Concentration at Voltage Manual Zero.
4.1.2.10.9 Oxygen Concentration at Voltage Manual Full Scale
This menu sets the oxygen concentration at which the analog voltage output is at maximum scale.
This setting may be set anywhere from 0.0 to 100.0% oxygen. The zero scale may be set higher
than the full scale for the Analog Voltage so that the scale may be inverted. Navigate from Analog
Output Setup menu to the Analog Output Calibration menu by typing "3" on the RS-232 terminal.
Then, type "7" to access the Voltage at Display Range Full Scale menu.
4.1.2.10.10
Oxygen Concentration at Current Manual Zero
This menu sets the oxygen concentration at which the analog current output is at minimum scale.
This option may be set anywhere from 0.0 to 100.0%. Navigate from Analog Output Setup menu
to the Analog Output Calibration menu by typing "3" on the RS-232 terminal. Then, type "8" to
access the Oxygen Concentration at Current Manual Zero menu.
4.1.2.10.11
Oxygen Concentration at Current Manual Full Scale
This menu sets the oxygen concentration at which the analog current output is a maximum scale.
This option maybe set anywhere from 0.0 to 100.0% oxygen. The zero scale may be set higher
than the full scale for the Analog Current so that the scale may be inverted. Navigate from Analog
Output Setup menu to the Analog Output Calibration menu by typing "3" on the RS-232 terminal.
Then, type "9" to access the Current at Display Range Full Scale menu.
4.1.2.10.12
Test Voltage Output Calibration
This menu may be used to test the Analog Voltage output. A digital multi-meter set to measure
volts must connected to TB3-pin5 and TB3-pin6 to verify the output. Any voltage between 0.0 and
1.0/5.0/10.0V may be entered at the prompt. Use the digital multi-meter to verify the Analog
Voltage output. Navigate from Analog Output Setup menu to the Analog Output Calibration menu
by typing "3" on the RS-232 terminal. Then, type "V" to access the Test Voltage Output Calibration
menu.
4.1.2.10.13
Test Current Output Calibration
This menu may be used to test the Analog Current output. A digital multi-meter set to measure
current must connected to TB3-pin6 and TB3-pin7 to verify the output. Any current between 0.0 to
20mA may be entered at the prompt. Use the digital multi-meter to verify the Analog Current
output. Navigate from Analog Output Setup menu to the Analog Output Calibration menu by typing
"3" on the RS-232 terminal. Then, type "I" to access the Test Current Output Calibration menu.
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018