STANDARD ACCESS level commands – continued:
TYPED COMMAND
1
2
3
4
5
Helpful Hint: Standard Level Access commands 1 through 5 can be used to manually cycle valves,
and controls for testing and troubleshooting the system and peripheral devices. Each time the
command is entered, the relay toggles its state.
Calibration solenoid valve is sticking, that valve may be cycled manually by pressing "3" on the PC
keyboard, and then pressing it again.
There are several data formats of the ASCII data dump available. They may be changed from
Standard Access level to suit the user's needs s follows:
TYPED COMMAND
S SERFMT=0
S SERFMT=1
S SERFMT=2
S SERFMT=3
S SERFMT=4
4.1.2.4.1 Disable RS 232 continuous output – SSERFMT=0
The factory default 1-second data-dump in Standard Level access can be disabled through the
RS-232 interface in Standard Access level. While this setting is active, the user must request
information by pressing the desired key according the STANDARD ACCESS level commands
chart.
Manual Part Number:
5-06-4900-53-0
Active Fault codes
Active Fault code descriptions
Long software version
Unit Serial Number
DESCRIPTION OF QUERIED FUNCTION
Manual O
Alarm relay override
2
Manual Fault relay override
Manual Remote Calibration relay override
Manual Inert Gas Control relay override
Cancel all manual relay overrides
DESCRIPTION OF QUERIED FUNCTION
Disables RS 232 continuous periodic data-dump
Enables RS 232 output in HUMAN READABLE format
Enables RS 232 output in MACHINE format w/o Checksum
Enables RS 232 output in MACHINE format w/Checksum
Enables RS 232 output in TAB DELIMITED (Excel) format
Revision Level: C
ECO 044952
If for example, the user suspects that the
Revision Date:
May 08, 2018
System Setup
Page
4-6