User Setup 2: O2 Alarm Relays Ascending Trip; User Setup 3: Analog Voltage Output Setting; User Setup 4: Analog Current Output Setting; User Setup 5: Display Range Setup - Bacharach Neutronics Mini-ICS Operation Manual

Oxygen analyzer / controller
Table of Contents

Advertisement

4.1.1.2

User Setup 2: O2 Alarm Relays Ascending Trip.

This parameter allows the user to set the O
Ascending trip is defined as; the relay is set to active when the Oxygen level is above the O
level set point. Descending trip is defined as; the relay is set to active when the Oxygen level is
below the O
Alarm level set point.
2
Valid Settings: 0 (Descending), 1 (Ascending).
4.1.1.3

User Setup 3: Analog Voltage Output Setting.

This parameter allows the user to set the Analog Output Voltage full scale to 1, 5, or 10 volts. Note
that the software settings must match the RA and RB jumper settings on the Main CPU PCB
(section 4.1.3).
Valid Settings: 0 (0-5V), 1 (0-10V), 2 (0-1V).
4.1.1.4

User Setup 4: Analog Current Output Setting.

This parameter allows the user to set the Analog Output Current range to either 0-20mA or 4-
20mA.
Valid Settings: 16 (4-20mA), 20 (0-20mA).
4.1.1.5

User Setup 5: Display Range Setup.

This parameter allows the user to map the scale of both the Analog Voltage and Current outputs,
based on the Oxygen concentration.
Valid Settings: 0 (0-1%), 1 (0-10%), 2 (0-25%), 3 (0-50%), 4(Auto-range), 5(Manual-Range)
4.1.1.6

User Setup 6: RS-232 Baud Rate.

This parameter allows the user to set the RS-232 communications baud rate.
Valid Settings: 0 (110bps), 1 (300bps), 2 (1200bps), 3 (2400bps), 4(4800bps), 5 (9600bps), 6
(19200bps), 7 (38400bps).
4.1.1.7

User Setup 7: Serial Output Format.

This parameter allows the user to set the RS-232 communications timed output format.
Valid Settings: 0 (Output on Request), 1 (Human Readable), 2 (Machine Code), 3 (Machine Code
with Checksum), 4(Tab Delimited).
4.1.1.8

User Setup 8: Sensor Disconnect Detection Test.

This parameter allows the user to enable or disable the sensor disconnect testing.
disconnect is used to determine whether or not there the sensor is connected to the Model Mini-
ICS analyzer.
Valid Settings: 0 (Disable Sensor Disconnect Detect), 1 (Enable Sensor Disconnect Detect).
4.1.1.9

User Setup 9: Minimum Analog Output Voltage Concentration

This parameter is only valid if Menu 5 is set to option 5 – Manual Range. This sets the oxygen
concentration at which the Analog Voltage is at minimum output.
Valid Settings: Any oxygen concentration between 0 to 100%
4.1.1.10

User Setup 10: Maximum Analog Output Voltage Concentration

This parameter is only valid if Menu 5 is set to option 5 – Manual Range. This sets the oxygen
concentration at which the Analog Voltage is at maximum output.
Manual Part Number:
5-06-4900-53-0
Alarm relay to trip either ascending or descending.
2
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
System Setup
Alarm
2
Sensor
Page
4-2