ILX Lightwave LDC-3916338 User Manual page 87

3.0 amp laser diode current source module
Hide thumbs Also See for LDC-3916338:
Table of Contents

Advertisement

¨ Front Panel
n Remote
The STATMENU:LINE1? query returns name of the measurement that is currently being displayed on line 1 of
the status menu.
SYNTAX DIAGRAM
PARAMETERS
None. The response is character data which represents the measurement that is currently being
displayed on line 1 of the selected channel's portion of the status menu.
POINTS OF
INTEREST
The possible responses are 1: "LDI" laser diode current.
The default condition of the status menu displays laser diode current ( LDI ) on line 1 and
photodiode current ( IPD ) on line 2.
EXAMPLES
"STATMENU:LINE1?"
displayed on line 1.
¨ Front Panel
n Remote
The STATMENU:LINE1: command path is used to get to the selected channel's status menu display selection
commands.
The following commands may be reached directly from the STATMENU:LINE1: command path.
STATMENU:LINE1:IPD
STATMENU:LINE1:LDI
STATMENU:LINE1:PPD
STATMENU:LINE1:VF
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
STATMENU
2: "IPD" photodiode current.
3: "PPD" optical power.
4: "VF" laser diode forward voltage.
-response:
"PPD", means that optical power is selected to be
PAGE 42
STATMENU:LINE1?
LINE1
?
STATMENU:LINE1:
CHPT 4

Advertisement

Table of Contents
loading

Table of Contents