ILX Lightwave LDC-3916338 User Manual page 52

3.0 amp laser diode current source module
Hide thumbs Also See for LDC-3916338:
Table of Contents

Advertisement

n Front Panel
n Remote
The LASer:CAL:ABORT command is used to exit a Laser calibration without changing the calibration values.
SYNTAX DIAGRAM
PARAMETERS
None.
POINTS OF
INTEREST
After this command is issued it is safe to run the LDC-3916338 Laser Diode Current Source with
the same calibration values used as before the LAS calibration mode was started.
This command is intended for use after a false or invalid calibration value is entered, or to exit
calibration mode before any values have been entered.
EXAMPLES
"Las:CAL:ABORT" -action: the LDC-3916338 Laser Diode Current Source exits calibration
mode.
n Front Panel
n Remote
The LASer:CAL:DEFAULT command is used to restore all of the module's LAS calibration default values.
SYNTAX DIAGRAM
PARAMETERS
None.
POINTS OF
INTEREST
This command should not be used except when normal calibration procedures and/or normal
operation of the LAS functions is not possible.
After this command is issued, the LAS module will not be calibrated. Each of the prescribed
calibration procedures must be redone. The default values are the starting point values which
exist before factory calibration is done.
EXAMPLES
"Las:CAL:Default" -action: the LDC-3916338 Laser Diode Current Source sets all of the LAS
calibration values to defaults on the selected module.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
LAS
:
:
LAS
PAGE 7
LASer:CAL:ABORT
:
CAL
ABORT
LASer:CAL:DEFAULT
:
CAL
DEFAULT
CHPT 4