ILX Lightwave LDC-3916338 User Manual page 81

3.0 amp laser diode current source module
Hide thumbs Also See for LDC-3916338:
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:STEP command specifies the amount the current setpoint will change in response to the LASer:INC
or LASer:DEC command.
SYNTAX DIAGRAM
:
PARAMETERS
An <nrf value> of the step amount, in the range 1.0 to 100.0. The resolution is 0.1.
POINTS OF
INTEREST
The step of 0.1 corresponds to 0.1 mA.
EXAMPLES
"Las:ldi 20; Las:Step 10.0; Las:Inc 1,1; Las:set:i?" -action: sets the selected channel's step to
10.0 mA, so the Las:set:i? query will return a value of 30.0 mA.
"LAS:STEP 10" -action: sets the selected channel's step size to 10.00 mA.
¨ Front Panel
n Remote
The LASer:STEP? query is used to read back the selected channel's STEP value. This value, in mA, is used by
the Las:Inc and Las:Dec commands to increment or decrement the laser current setpoint.
SYNTAX DIAGRAM
PARAMETERS
None. The response is an <nrf value> of the selected channel's step amount.
POINTS OF
INTEREST
The step of 1 corresponds to 1.0 mA.
EXAMPLES
"Las:Step?" -response: 1.0 means the selected channel's step size is 1.0 mA.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
LAS
STEP
:
:
LAS
PAGE 36
LASer:STEP
<white
<nrf
space>
value>
LASer:STEP?
?
STEP
CHPT 4