n Front Panel
n Remote
The LASer:CAL:MEAS command is used to enter the Laser calibration values for the selected channel.
SYNTAX DIAGRAM
:
LAS
PARAMETERS
An <nrf value> which represents the calibration value expected for the present LAS calibration
state.
POINTS OF
INTEREST
In remote operation, the LAS:CAL:STAT? query may be used to determine if the LDC-3916338
Laser Diode Current Source is ready for the user to enter a value via the LAS:CAL:MEAS
command.
EXAMPLES
"Las:CAL:MEAS 4.95" -action: the LDC-3916338 Laser Diode Current Source enters the value
4.95 for calibration value of the present calibration mode.
¨Front Panel
n Remote
The LASer:CAL:STATus? query is to determine whether the unit is ready for a calibration value and whether it
is in calibration mode.
SYNTAX DIAGRAM
PARAMETERS
None. The response will be an <nrf value>; 0 = Not Ready for Measurement, 1 = Ready for
Measurement, 2 = Calibration Completed, and 3 = Calibration Aborted.
POINTS OF
INTEREST
If the calibration is aborted, the status value will remain "3" until the calibration is started again.
EXAMPLES
"LAS:CAL:STAT?" -response: 1, means the LDC-3916338 Laser Diode Current Source is ready
for a calibration measurement to be entered via the "LAS:CAL:MEAS" command.
"LAS:CAL:STAT?" -response: 2, means the LDC-3916338 Laser Diode Current Source has
completed the LAS calibration.
"LAs:Cal:Stat?" -response: 0, means the LDC-3916338 Laser Diode Current Source is not yet
ready for the LAS calibration measurement value to be entered.