ILX Lightwave LDC-3916338 User Manual page 60

3.0 amp laser diode current source module
Hide thumbs Also See for LDC-3916338:
Table of Contents

Advertisement

¨ Front Panel
n Remote
The LASer:ENABle:EVEnt command sets the selected channel's Laser Event Status Enable register. The bits
in this register determine which bits in the Event Status register are summarized in the 3916 or 3908 Status
Byte.
SYNTAX DIAGRAM
:
LAS
PARAMETERS
An <nrf value> whose sum represents the bits which are enabled:
1 - Laser Current Limit Changed State
2 - Laser Voltage Limit Changed State
4 - N/A
8 - Power Limit Changed State
16 - Interlock Changed State
32 - N/A
64 - N/A
128 - Open Circuit Changed State
POINTS OF
INTEREST
The enabled Laser events for the selected channel can be read by using the
LASer:ENABle:EVEnt? query. The selected channel's Laser event status can be monitored by
the LASer:EVEnt? query.
EXAMPLES
"LAS:ENAB:EVENT 136" -action: only the Open Circuit and Power Limit events will be
summarized in the status byte.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
:
ENAB
PAGE 15
LASer:ENABle:EVEnt
<white
<nrf
EVE
value>
space>
256 - Output is Off Changed State
512 – Output Changed Tolerance State
1024 - Output is On Changed State
2048 - N/A
4096 - N/A
8192 - N/A
16384 - N/A
32768 - N/A