n Front Panel
n Remote
The LASer:LDI command sets the laser current setpoint for the selected channel.
SYNTAX DIAGRAM
:
PARAMETERS
An <nrf value> which represents the (laser) output current, in mA.
POINTS OF
INTEREST
The setpoint is the same for both low and high bandwidth output modes.
EXAMPLES
"Las:LDI 400" -action: sets the selected channel's laser output current to 400.00 mA.
n Front Panel
n Remote
The LASer:LDI? query returns the value of the measured laser current for the selected channel.
SYNTAX DIAGRAM
PARAMETERS
None. The response is an <nrf value>.
POINTS OF
INTEREST
Response is the selected channel's measured laser output current, regardless of LAS control
mode.
This measurement is updated approximately once every 600 msec.
EXAMPLES
"LAS:ldi?" -response: 30.0, means the selected channel's measured laser output current is 30.0
mA.
"Laser:LDI?" -response: 149.6, means the selected channel's measured laser output current is
149.6 mA.
COMMAND REFERENCE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
LDI
LAS
:
:
LAS
PAGE 21
LASer:LDI
<white
<nrf
space>
value>
LASer:LDI?
?
LDI
CHPT 4