Jeio tech Lab companion SI-300 Operating Instruction page 48

Shaking incubator
Table of Contents

Advertisement

(2) Electricity
Phenomenon
Even if Power Switch
ON, OFF promptly..
Even if Power Switch
ON, an indication
does not work in
Display Panel.
If temperature does
not rise
If a machine stops
operation by Over Temp.
If the unit stops
operation
by Door Limit Switch
Stops total control
without an indication
except the above
statement or button
operation.
Check point
Leak.
A power plug being
Disconnected from an outlet.
Fuse Problem
Confirms whether Temp Led
is turned on
1 O/T LED is turned on
with red,
2. confirms whether a
Beep sound rings successively
1. Door Led is turned on
with red
2.
Confirms
whether
Beep sound rings
1.Being affected by
strong Noise in a power
line
2. confirms whether the
high frequency welder
which strong high
frequency Noise occurs in
around appliances, mass
SCR Controller became
installation at hand
receive correct
acceptance.
Settlement
Stops a driving immediately and contacts
a reseller
Leaves Power Switch OFF, and please put a
power plug in an outlet.
Exchange the fuse positioned backside.
If Temp Led is out, press Start/Stop
button.
1 If attributes red Knob of Over Temp.
Limit to a (-) driver clockwise and
presses one time of Start/Stop button
after currently adjusting than
temperature (PV) higher than 15%, red of
O/T Led disappears, and a Beep sound
disappears.
2. While Run LED and Heater LED are on if
they press one time of Start/Stop button
more, a machine operates again.
1 Check whether Door LED is flickering
due to opened door
a
2. Confirms whether light reaches RUN LED
after having pressed one time of
START/STOP button.
Moves equipment to be the cause of, or a
change holds an Incubator installation
situation.