9. Causes of malfunction and its repairs
1. When the unit stops running
Troubleshooting
When the unit stops running
2. Settlement Details
(1) The refrigeration
Phenomenon
The
temperature
decrease.
The temp is unstable.
Discrepancy between Displayed
temp and present temp.
Page 46
Check the main power.
Check the Fuse..
Check the RUN Led display is off.
If the unit is off, Press the START/STOP button to
operate..
Check the Electric break.
Check Point
Check
the
illuminated
Check
the
illuminated
doesn' t
Check
the
Condenser.
Check the inner circulation
fan speed is too low..
Check the PID value.
Check the Blower speed.
Check the sample is too much
inside the chamber.
Check the revision value.
Check point and Settlement
TEMP
LED
is
Press the START/STOP Button.
W/ON
LED
is
Release the Timer mode
dust
on
the
Clean the filter opened the
Condenser cover.
Adjust the fan speed high.
Operates the Auto Tuning.
(Reference page 27)
Adjust the Blower fan speed
high .(Reference page 24)
Remove the sample to flow the
air.
Revise
between
present temp..
(Reference :Page 22)
Settlement
the
discrepancy
Displayed
temp
and