Problem
Fault 559
Fault 578
Process-related faults and messages
Problem
Fit a new coarse filter!
Fit a new HEPA filter!
Drying during programme de-
activated
Wrong code entered
Cause and remedy
There is a problem with the process documentation
interface. The machine has detected a module for an
Ethernet interface, but only a serial interface (RS232) is
activated in the controls.
Deactivate the RS232 interface:
Open the menu for configuring the interface
settings
/
Interface
Wait approx. 90 seconds. The Ethernet module XKM
3000 L Med needs this time for initialisation. It may be
necessary to reconfigure the interface.
Or
Replace the Ethernet module XKM 3000 L Med with a
XKM RS232 10 Med module to set up a serial interface.
The peak load cut-out has lasted longer than 3 hours.
Have your electrical system and your energy
management system tested by a suitably qualified
person.
Cause and remedy
The maximum permissible operating hours for the coarse
filter have been reached.
Replace the coarse filter with a new one.
Reset the operating hours counter for the coarse filter.
The maximum permissible operating hours for the HEPA
filter have been reached.
Replace the HEPA filter with a new one.
Reset the operating hours counter for the HEPA filter.
Drying cannot be selected at the start of a programme
because drying is not available for the selected
programme.
Start the programme without drying.
Or:
Have the drying parameters for this programme adjusted
by the Miele Customer Service Department.
The PIN code entered is not the same as the code saved.
Enter the PIN code again.
Report the loss of the PIN code to the Miele Customer
Service Department.
Problem solving guide
and then select
Ethernet
Further
.
141