Filter Change; Changing The Coarse Filter - Miele PG 8583 CD Operating Instructions Manual

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Maintenance
Changing the
coarse filter
128

Filter change

The air filter for the internal drying unit has a limited lifespan and has
to be replaced at regular intervals, e.g. when the following message
appears:
Open the drawer in the side unit.
Release the filter grille and swing it downwards to open. The grille
can also be removed completely.
Change the coarse filter. The soft side of the filter must be to the
front.
Replace the filter grille and close the side unit drawer.
Whenever the coarse filter is replaced the operating hours counter
must be reset (see "Resetting the operating hours counter").
Fit a new HEPA filter!