Miele PG 8583 CD Operating Instructions Manual page 125

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Use a pointed object to push particles into the spray arm.
Rinse the spray arm thoroughly under running water.
 Do not allow any magnetic objects or load items to attach to the
magnets on the spray arms.
Any metallic objects on the magnets can cause a false reading of
spray arm rotation.
Remove all metal objects from the magnets.
Check the spray arm bearings for visible signs of wear.
Visible wear on the bearings can adversely affect the long-term
functioning of the spray arms.
In this case, contact the Miele Customer Service Department.
Replace the spray arms after cleaning.
Make sure the spray arms can rotate easily after they have been
fitted.
The spray arms and baskets each have a number e.g. 03, which is
also embossed on the water supply pipes near the bayonet fittings.
When refitting, ensure that the numbers on the spray arms
correspond with the numbers on the water supply pipes.
Maintenance
125

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pg 8583 cd ww adp cmPg 8583 cd ww ad cm oil

Table of Contents