Miele PG 8583 CD Operating Instructions Manual page 133

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Activating a test
programme
Test programmes are valid for only one programme sequence each
time. A test programme must be selected again for further tests.
Open the menu as follows:
 button
Further settings
     
Test programme
         
Test programme
No
Laboratory
No
-
The menu is exited without selecting a programme.
Laboratory
-
Activates the Laboratory test programme
-
Validation
Activates the Validation test programme
Select an option using the  and  arrow buttons.
Press OK to activate the test programme for the next programme
start.
You can now start the performance test.
Select and start a programme using the programme selection
buttons or via the programme list.
The programme will be identified in the bottom line as
programme
during the programme sequence.
If you want to deactivate the test programme before the performance
test you need to go to the next menu level up and select the
option.
Maintenance
Test
No
133

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pg 8583 cd ww adp cmPg 8583 cd ww ad cm oil

Table of Contents